Liquidus Temperature: Assessing Standard Glasses for Furnace Calibration

Abstract

Standard glasses are used to ensure accurate liquidus temperature measurements and furnace temperatures. The supply of the standard glasses used at Pacific Northwest National Laboratory (Standard Reference Material 773 [SRM-773]) has been depleted. Replacement standard glasses for furnace temperature verification and liquidus temperature measurement validation were fabricated and tested. Two glasses-ARG-1 and AmCm2-19-were fabricated in roughly 5 kg batches. The liquidus temperature of the glasses was measured and found to be consistent with those reported in literature.