We report the use of laser-induced fluorescence (LIF) of laser ablation plumes for standoff applications. The standoff analysis of Al species, as major and minor species in samples, is performed in a nanosecond laser-produced plasma created at a distance ~10 m. The LIF analysis is performed by resonantly exciting an Al transition at 394.4 nm using a continuous wave (cw) tunable laser and by collecting the direct-line fluorescence signal at 396.15 nm. The spectral resolution of LIF is obtained by scanning the cw tunable LIF laser across the selected Al transition. Our results highlight that LIF provides enhanced signal intensity, emission persistence, and spectral resolution when compared to thermally-excited emission, and these are crucial considerations for using laser-produced plasma for standoff isotopic analysis.
Revised: March 22, 2019 |
Published: February 26, 2018
Citation
Harilal S.S., B.E. Brumfield, and M.C. Phillips. 2018.Standoff analysis of laser-produced plasmas using laser-induced fluorescence.Optics Letters 43, no. 5:1055-1058.PNNL-SA-130891.doi:10.1364/OL.43.001055