We have designed and implemented a Cray XD-1-based sim- ulation of data capture and signal processing for an ad- vanced Ion Mobility mass spectrometer (Hadamard trans- form Ion Mobility). Our simulation is a hybrid application that uses both an FPGA component and a CPU-based soft- ware component to simulate Ion Mobility mass spectrome- try data processing. The FPGA component includes data capture and accumulation, as well as a more sophisticated deconvolution algorithm based on a PNNL-developed en- hancement to standard Hadamard transform Ion Mobility spectrometry. The software portion is in charge of stream- ing data to the FPGA and collecting results. We expect the computational and memory addressing logic of the FPGA component to be portable to an instrument-attached FPGA board that can be interfaced with a Hadamard transform Ion Mobility mass spectrometer.
Revised: September 30, 2010 |
Published: November 3, 2007
Citation
ChavarrÃa-Miranda D., B.H. Clowers, G.A. Anderson, and M.E. Belov. 2007.Simulating data processing for an Advanced Ion Mobility Mass Spectrometer. In Proceedings of the 1st international workshop on High-performance reconfigurable computing technology and applications (HPRCTA'07): held in conjunction with SC07, NOV 10-16, 2007, Reno Nevada, 21-29. New York, New York:Association for Computing Machinery.PNNL-SA-57104.