We report on a new electrodynamic ion funnel that operates at a pressure of 30 Torr with no loss of ion transmission. The enhanced performance compared to previous ion funnel designs optimized for pressures of
Revised: October 1, 2015 |
Published: September 1, 2006
Citation
Ibrahim Y.M., K. Tang, A.V. Tolmachev, A.A. Shvartsburg, and R.D. Smith. 2006.Improving Mass Spectrometer Sensitivity Using a High-Pressure Electrodynamic Ion Funnel Interface.Journal of the American Society for Mass Spectrometry 17, no. 9:1299-1305.PNNL-SA-49272.doi:10.1016/j.jasms.2006.06.005