Data-dependent selective external ion ejection with improved resolution is demonstrated with a 3.5 tesla FTICR instrument employing DREAMS (Dynamic Range Enhancement Applied to Mass Spectrometry) technology. To correct for the fringing rf-field aberrations each rod of the selection quadrupole has been segmented into three sections, so that ion excitation and ejection was performed by applying auxiliary rf-only waveforms in the region of the middle segments. Two different modes of external ion trapping and ejection were studied with the mixtures of model peptides and a tryptic digest of bovine serum albumin. A mass resolution of about 100 had been attained for rf-only dipolar ejection in a quadrupole operating at a Mathieu parameter q of ~0.45. LC-ESI-DREAMS-FTICR analysis of a 0.1 mg/mL solution of bovine serum albumin digest resulted in detection of 82 unique tryptic peptides with mass measurement errors lower than 5 ppm, providing 100 % sequence coverage of the protein.