An experimental study of sub-ambient pressure electrosprays is reported. The pressure domain that affords stable electrospray operation appears to be limited by the vapor pressure of the liquid. The voltage driving the electrospray is shown to have a logarithmic dependence on the pressure. This scaling amends the relationship currently in use to calculate the electric field at the tip of the meniscus of an electrified liquid
Revised: June 23, 2010 |
Published: November 2, 2009
Citation
Marginean I., J.S. Page, R.T. Kelly, K. Tang, and R.D. Smith. 2009.Effect of pressure on electrospray characteristics.Applied Physics Letters 95, no. 18:Article Number: 184103.PNNL-SA-65737.