Hybrid two-stage mass spectrometers capable of both highly accurate mass measurement and MS/MS fragmentation have become widely available in recent years and have allowed for sig-nificantly better discrimination between true and false MS/MS pep-tide identifications by applying relatively narrow windows for maxi-mum allowable deviations for parent ion mass measurements. To fully gain the advantage of highly accurate parent ion mass meas-urements, it is important to limit systematic mass measurement errors. The DtaRefinery software tool can correct systematic errors in parent ion masses by reading a set of fragmentation spectra, searching for MS/MS peptide identifications, then fitting a model that can estimate systematic errors, and removing them. This results in a new fragmentation spectrum file with updated parent ion masses.
Revised: March 11, 2010 |
Published: December 16, 2009
Citation
Petyuk V.A., A.M. Mayampurath, M.E. Monroe, A.D. Polpitiya, S.O. Purvine, G.A. Anderson, and D.G. Camp, et al. 2009.DtaRefinery: a software tool for elimination of systematic errors from parent ion mass measurements in tandem mass spectra datasets.Molecular & Cellular Proteomics. MCP 9.3.PNNL-SA-62317.doi:10.1074/mcp.M900217-MCP200