November 1, 2004
Journal Article

Tailored Noise Waveform/ Collision-Induced Dissociation of Ions Stored in a Linear Ion Trap Combined with Liquid Chromatography/Fourier Transform Ion Cyclotron Resonance Mass Spectrometry

Abstract

A new collision-induced dissociation (CID) technique based on broadband tailored noise waveform (TNW) excitation of ions stored in a linear ion trap has been developed. In comparison with the conventional sustained off-resonance irradiation (SORI) CID method commonly used in Fourier transform ion cyclotron resonance mass spectrometry, this MS/MS technique increases throughput by eliminating the long pump-down delay associated with gas introduction into the high vacuum ICR cell region. In addition, the TNW-CID method speeds spectrum acquisition since it does not require Fourier transformation, calculation of resonant frequencies and generation of the excitation waveforms. We demonstrate TNW-CID coupled with on-line capillary reverse phase liquid chromatography separations for identification of peptides. The experimental results are compared with data obtained using conventional quadrupole ion trap MS/MS and SORI-CID MS/MS in an ICR cell.

Revised: September 23, 2015 | Published: November 1, 2004

Citation

Vilkov A.N., B. Bogdanov, L. Pasa-Tolic, D.C. Prior, G.A. Anderson, C.D. Masselon, and R.J. Moore, et al. 2004. Tailored Noise Waveform/ Collision-Induced Dissociation of Ions Stored in a Linear Ion Trap Combined with Liquid Chromatography/Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. Rapid Communications in Mass Spectrometry 18, no. 22:2682-2690. PNNL-SA-42697.