Abstract

A new sensor was developed for simultaneous detection of cadmium (Cd2+), copper (Cu2+), and lead (Pb2+), based on the voltammetric response at a carbon paste electrode modified with carbamoylphosphonic acid (acetamide phosphonic acid) self-assembled monolayer on mesoporous silica (Ac-Phos SAMMS). The adsorptive stripping voltammetry technique involves preconcentration of the metal ions onto Ac-Phos SAMMS under an open circuit, then electrolysis of the preconcentrated species, followed by a square wave potential sweep towards positive values. Factors affecting the preconcentration process were investigated. The voltammetric responses increased linearly with the preconcentration time from 1 to 30 minutes or with metal ion concentrations ranging from 10 to 200 ppb. The responses also evolved in the same fashion as adsorption isotherm in the pH range of 2-6. The metal detection limits were 10 ppb after 2 minutes preconcentration and improved to 0.5 ppb after 20 minutes preconcentration.

Revised: February 20, 2004 | Published: January 30, 2004

Citation

Yantasee W., Y. Lin, G.E. Fryxell, and B.J. Busche. 2004. Simultaneous Detection of Cadmium, Copper, and Lead using A Carbon Paste Electrode Modified with Carbamoylphosphonic Acid Self-Assembled Monolayer on Mesoporous Silica (SAMMS). Analytica Chimica Acta 502, no. 2:207-212. PNNL-SA-39324.