The ability to conduct automated trace radionuclide analysis at or near the sample collection point would provide a valuable tool for emergency response, nuclear forensics and environmental monitoring. Pacific Northwest National Laboratory is developing systems for this purpose based on dual gamma-ray spectrometers, e.g. NaI(TI) or HPGe, combined with thin organic scintillator sensors to detect light charged particles. Translating the coincident signatures recorded by these systems, which include , and , into the concentration of detectable radionuclides in the sample requires generalized multicoincidence analysis tools. The development and validation of the Coincidence Lookup Library, which currently contains the probabilities of single and coincidence signatures from more than 420 isotopes, is described. Also discussed is a method to calculate the probability of observing a coincidence signature which incorporates true coincidence summing effects. These effects are particularly important for high-geometric-efficiency detection systems. Finally, a process for validating the integrated analysis software package is demonstrated using GEANT 4 simulations of the prototype detector systems.
Revised: June 23, 2011 |
Published: October 1, 2005
Citation
Warren G.A., L.E. Smith, C.E. Aalseth, J.E. Ellis, A.B. Valsan, and W. Mengesha. 2005.Generalized Multicoincidence Analysis Methods. In 2004 IEEE Nuclear Science Symposium Conference Record, 1, 345-349. Piscataway, New Jersey:IEEE.PNNL-SA-42995.doi:10.1109/NSSMIC.2004.1462210