A radiation detection system consisting of two cerium doped lanthanum bromide (LaBr3:Ce) scintillation
detectors in a gamma-gamma coincidence configuration has been used to demonstrate the advantages that
coincident detection provides relative to a single detector, and the advantages that LaBr3:Ce detectors provide
relative to high purity germanium (HPGe) detectors. Signal to noise ratios of select photopeak pairs for these
detectors have been compared to high-purity germanium (HPGe) detectors in both single and coincident detector
configurations in order to quantify the performance of each detector configuration. The efficiency and energy
resolution of LaBr3:Ce detectors have been determined and compared to HPGe detectors. Coincident gamma-ray
pairs from the radionuclides 152Eu and 133Ba have been identified in a sample that is dominated by 137Cs.
Gamma-gamma coincidence successfully reduced the Compton continuum from the large 137Cs peak, revealed
several coincident gamma energies characteristic of these nuclides, and improved the signal-to-noise ratio
relative to single detector measurements. LaBr3:Ce detectors performed at count rates multiple times higher than
can be achieved with HPGe detectors. The standard background spectrum consisting of peaks associated with
transitions within the LaBr3:Ce crystal has also been significantly reduced. It is shown that LaBr3:Ce detectors
have the unique capability to perform gamma-gamma coincidence measurements in very high count rate
scenarios, which can potentially benefit nuclear safeguards in situ measurements of spent nuclear fuel.
Revised: November 6, 2017 |
Published: January 15, 2017
Citation
Drescher A., M. Yoho, S. Landsberger, M. Durbin, S. Biegalski, D.E. Meier, and J.M. Schwantes. 2017.Gamma-gamma coincidence performance of LaBr3:Ce scintillation detectors vs HPGe detectors in high count-rate scenarios.Applied Radiation and Isotopes 122.PNNL-SA-129947.doi:10.1016/j.apradiso.2017.01.012