The kinetics of charge reduction and desorption of different species produced by soft-landing of mass-selected ions was studied using in situ secondary ion mass spectrometry (SIMS) in a Fourier transform ion cyclotron resonance mass spectrometer (FT-ICR MS). The improved SIMS capability described in this work utilizes an in-line 8 keV Cs+ ion gun and allows us to interrogate the surface both during the ion deposition and after the deposition is terminated. As a model system doubly protonated ions of Gramicidin S were deposited onto a fluorinated self-assembled monolayer (FSAM) surface. Our results demonstrate for the first time that various peptide-related peaks in FT-ICR SIMS spectra follow very different kinetics. We obtained unique kinetics signatures for doubly protonated, singly protonated and neutral peptides retained on the surface and followed their evolution as a function of time. The experimental results are in excellent agreement with a kinetic model that takes into account charge reduction and thermal desorption of different species from the surface.
Revised: May 8, 2008 |
Published: December 13, 2007
Citation
Hadjar O., J.H. Futrell, and J. Laskin. 2007.First Observation of Charge Reduction and Desorption Kinetics of Multiply Protonated Peptides Soft Landed onto Self-Assembled Monolayer Surfaces.Journal of Physical Chemistry C 111, no. 49:18220-18225.PNNL-SA-56122.doi:10.1021/jp075293y