Damage-free analysis of biological materials by vibrational spectroscopy in the EM

Abstract

Atomic-resolution compositional analysis in an electron microscope using electron energy loss spectroscopy (EELS) or energy-dispersive X-ray spectroscopy (EDXS) requires doses of 106 to 1010 e- / Å2. Doses of this order completely destroy biological samples. Compositional information on biological samples can nevertheless be obtained from frozen hydrated biological samples with worse spatial resolution at much lower doses [1], or at higher doses on freeze-dried samples [2, 3]. However, the accompanying radiation damage is severely limiting, and EELS and EDXS microanalysis have not been adopted widely in biological electron microscopy. Vibrational spectroscopy in the electron microscope, introduced 6 years ago [4], can overcome radiation damage by acquiring spectra with the beam positioned a few tens of nm away from a beam-sensitive sample, in the so-called aloof scattering geometry. Spectra bearing a close correspondence to FTIR data are then obtained, and the sample suffers essentially no radiation damage [4-9]. At 5 meV energy resolution, which is now readily attainable, aloof vibrational spectra recorded in this way are able to detect even subtle effects such as isotopic substitution of 13C for 12C [6]. Fig. 1 shows background-subtracted vibrational spectra acquired from the sugar trehalose, used by bacteria for energy storage, and from bovine serum albumin (BSA), an animal protein. Features near 400 meV are due to vibrations of hydrogen atoms that typically lie on the outside of the molecules, and these can change as a function of the environment the molecule is in. Features near 200 meV and below are typically due to the vibrations of heavier atoms such as oxygen, carbon and nitrogen, often on the inside of the molecule. They are typically independent of the molecule’s environment, and can be used for detecting the presence of different biological molecules in complex samples. Spectra acquired in aloof geometry from outside the sample are not useful for general biological samples, in which the interesting structures occur away from the sample’s edge. However, similar spectra can be obtained when the electron beam is placed on a beam-sensitive sample and proceeds to drill a hole [7], at doses of around 103 e- /Å. This arises because the drilled hole is typically 10 nm, the material between the holes appears to suffer little radiation damage [9]. In summary, vibrational spectroscopy in the electron microscope promises to probe biological materials in a way that is similar to infrared spectroscopy, at a spatial resolution of about 30-50 nm, while circumventing radiation damage. We are now focused on developing this technique fully, using a side- entry cryotransfer holder in a Nion HERMES electron microscope equipped with a fast and efficient EELS direct detector [10], and readily capable of 5 meV energy resolution. The results of these investigations will be reported at the meeting. References [1] R.D. Leapman and S. Sun, Ultramicroscopy 59 (1995) 71. [2] J. Feng, A.V. Somlyo and A.P. Somplyo, J. of Micr. 215 (2004) 92. [3] A.V. Somlyo, H. Shuman and A.P. Somlyo, Methods Enzymol. 172 (1989) 203. [4] O.L. Krivanek et al., Nature 514 (2014) 209; T. Miata et al., Mic

Published: September 19, 2024

Citation

Krivanek O., A. Dohnalkova, Z. Kochovski, B. Haas, J. Muller, N. Dellby, and M. Hotz, et al. 2020. Damage-free analysis of biological materials by vibrational spectroscopy in the EM. Microscopy and Microanalysis 26, no. S2:108–110. PNNL-SA-155545. doi:10.1017/S1431927620013409