The necessity to monitor international commercial transportation for illicit nuclear materials resulted in the installation of many nuclear radiation detection systems in Portal Monitors. These were mainly gross counters which alarmed at any indication of high radioactivity in the shipment, the vehicle or even the driver. The innocent alarm rate, due to legal shipments of sources and NORM, or medical isotopes in patients, caused interruptions and delays in commerce while the legality of the shipment was verified. To overcome this difficulty, Department of Homeland Security (DHS) supported the writing of the ANSI N42.38 standard (Performance Criteria for Spectroscopy-Based Portal Monitors used for Homeland Security) to define the performance of a Portal Monitor with nuclide identification capabilities, called a Spectroscopy Portal Monitor. This standard defines detection levels and response characteristics for the system for energies from 25 keV to3. MeV. To accomplish the necessary performance, several different HPGe detector configurations were modeled using MCNP for the horizontal field of view (FOV) and vertical linearity of response over the detection zone of 5 meters by 4.5 meters for 661 keV as representative of the expected nuclides of interest. The configuration with the best result was built and tested. The results for the FOV as a function of energy and the linearity show good agreement with the model and performance exceeding the requirements of N42.38.
Revised: September 4, 2008 |
Published: May 15, 2008
Citation
Keyser R.M., W.K. Hensley, T.R. Twomey, D.L. UPP, and D.L. UPP. 2008.Comparison of MCNP and Experimental Measurements for an HPGe-based Spectroscopy Portal Monitor.Journal of Radioanalytical and Nuclear Chemistry 276, no. 2:399-405.PNNL-SA-60901.