In this paper, a vacuum compatible microfluidic device, System for Analysis at the Liquid Vacuum Interface (SALVI), is integrated to hard x-ray absorption spectroscopy (XAS) to obtain the local structure of K3[Fe(CN)6] in aqueous solutions with three concentrations of 0.5 M, 0.05 M, and 0.005 M. The solutions were sealed in a microchannel of 500 µm wide and 300 µm deep in a portable microfluidic device. The Fe K-edge x-ray absorption spectra show that the complex in water is Fe(III). The complex is present with octahedral geometry coordinated with 6 C atoms in the first shell with a distance of ~1.92 Å and 6 N atoms in the second shell with a distance of ~3.10 Å. Varying the concentration has no observable influence on the structure of K3[Fe(CN)6]. Our results demonstrate the feasibility of using microfluidic based liquid cells in large synchrotron facilities and it is a viable approach to enable multifaceted measurements of liquids in the future.
Revised: April 30, 2020 |
Published: April 11, 2018
Citation
Zheng J., W. Zhang, F. Wang, and X. Yu. 2018.Enabling liquid solvent structure analysis using hard x-ray absorption spectroscopy with a transferrable microfluidic reactor.Journal of Physics: Condensed Matter 30, no. 18:Article No. 18LT01.PNNL-SA-131647.doi:10.1088/1361-648X/aab87f