A rapid, simple and sensitive electrochemical assay of horseradish peroxidase (HRP) performed on disposable screen-printed carbon electrode was developed. HRP activities were monitored by square-wave voltammetric (SWV) measuring the electroactive enzymatic product in the presence of o-aminophenol and hydrogen peroxide substrate solution. SWV analysis demonstrated a greater sensitivity and shorter analysis time than the widely used amperometric and differential-pulsed voltammetric methods. The voltammetric characteristics of substrate and enzymatic product as well as the parameters of SWV analysis were optimized. Under optimized conditions, a linear response for HRP from 0.003 - 0.1 U/mL and a detection limit of 0.002 U/mL (1.25×10-15 mol in 25 µL) were obtained with a good precision (RSD = 8%; n = 6). This rapid and sensitive HRP assay with microliters-assay volume could be readily integrated to portable devices and point-of-care (POC) diagnosis applications.
Revised: August 6, 2010 |
Published: September 10, 2008
Citation
Lee A., G. Liu, C. Heng, S. Tan, T. Lim, and Y. Lin. 2008.Sensitive electrochemical detection of horseradish peroxidase at disposable screen-printed carbon electrode.Electroanalysis 20, no. 18:2040-2046.PNNL-SA-61143.