A swept-ECQCL is used for broadband IR spectroscopy of isotopic mixtures of CH3OH, CH3OD, CH3CH2OH, and CH3CH2OD in a static gas cell over a wavelength range of 9.5 to 10.4 µm. A weighted least squares fitting approach with quantitative library spectra illustrates that significant spectral congestion does not negatively impact the ability for in situ quantification of large isotopic species in a mixture. The noise equivalent concentrations for CH3OH, CH3OD, CH3CH2OH, and CH3CH2OD are 19 ppbv x m, 28 ppbv x m, 450 ppbv x m, and 330 ppbv x m respectively for a 50 second integration time. Based on the observed NECs, isotopic precisions of 0.07‰ and 0.79‰ for a 50 s integration time are calculated for measurements of the [MeOD]/[MeOH] and [EtOD]/[EtOH] isotope ratios , respectively, for the species concentrations in the gas cell.
Revised: July 31, 2017 |
Published: July 7, 2017
Citation
Brumfield B.E., and M.C. Phillips. 2017.Quantitative isotopic measurements of gas-phase alcohol mixtures using a broadly tunable swept external cavity quantum cascade laser.The Analyst 142, no. 13:2354-2362.PNNL-SA-123788.doi:10.1039/c7an00223h