A filter pillar-array microstructure was coupled with a pneumatic micro-valve to fabricate a reusable miniaturized beads-trapping/releasing flow cell, in which trapping and releasing beads can be conveniently realized by switching the micro-valve. This miniaturized device was suitable to construct automatic fluidic system for “renewable surface analysis”. The renewable surface strategy based on pneumatic micro-valve enabled capture of beads in beads chamber prior to each assay, and release of the used beads after the assay. Chemiluminescent competitive immunoassay of 3,5,6-trichloropyridinol (TCP) was performed as a model to demonstrate the application potential of this reusable miniaturized flow cell. The whole fluidic assay process including beads trapping, immuno-binding, beads washing, beads releasing and signal collection could be completed in 10 min. Immunoassay of TCP using this miniaturized device showed a linear range of 0.20-70 ng/mL with a limit of detection of 0.080 ng/mL. The device had been successfully used for detection of TCP spiked in rat serum with average recovery of 97%. This investigation provides a rapid, sensitive, reusable, low-cost and automatic miniaturized device for solid-phase biochemical analysis for various purposes.
Revised: April 15, 2011 |
Published: April 1, 2011
Citation
Fu Z., G. Shao, J. Wang, D. Lu, W. Wang, and Y. Lin. 2011.Microfabricated Renewable Beads-Trapping/Releasing Flow Cell for Rapid Antigen-Antibody Reaction in Chemiluminescent Immunoassay.Analytical Chemistry 83, no. 7:2685-2690.PNNL-SA-76859.