March 11, 2006
Report

Calculation of Minimum-Detectable-Concentration Levels of Radioxenon Isotopes Using the PNNL ARSA System

Abstract

Measurement of xenon fission product isotopes is a key element in the global network being established to monitor the Comprehensive Nuclear-Test-Ban Treaty. The automated Radio-xenon Analyzer/Sampler (ARSA), built by Pacific Northwest National Laboratory, can detect 131mXe, 133mXe, 133Xe, and 135Xe via a beta-gamma counting system. Due to the variable background and sources of these four radio-xenon isotopes, it is important to have as sensitive a detection system as possible and to quantify the Minimum-Detectable-Concentrations (MDC) that such a system will be able to detect to preclude false negative and false positive results. From data obtained from IAR in Germany MDC values for 133Xe were well below the 1 mBq/SCMA as required by the PTS for the Comprehensive Test BAn Treaty [WGB TL-11,1999].

Revised: November 8, 2007 | Published: March 11, 2006

Citation

McIntyre J.I., T.W. Bowyer, and P.L. Reeder. 2006. Calculation of Minimum-Detectable-Concentration Levels of Radioxenon Isotopes Using the PNNL ARSA System Richland, WA: Pacific Northwest National Laboratory.